Determination of boron in the waters of Troia by inductively coupled plasma-atomic emission spectrometry (ICP-AES)

Yükleniyor...
Küçük Resim

Tarih

2005

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Serbian Chemical Soc

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

The inductively coupled plasma-atomic emission spectrometry method (ICP-EAS) is proposed for the determination of boron in Troia water sources. In this study, boron samples were taken from 25 different locations during November 2002 and May 2003. ne 249.773 nm boron emmision line was not affected by the sample matrix. The boron content was determined without further pretreatment of samples proposed ICP-AES method. An analysis of a given sample was completed in about 15 min. The determined boron concentrations in surface and ground water sources were between 0.00444-1.11200 mg l(-1). In order to compare the. results obtained by the proposed method, selected water samples were also analysed titrimetically. Pearson correlation coefficient analysis showed that there is a close relationship between the ICP-AES and the titrimetric method and the determination coefficient (R-2) was calculated to be 91.34 %.

Açıklama

Anahtar Kelimeler

boron, inductively coupled plasma-atomic emission spectroscopy, titrimetry, water analysis

Kaynak

Journal of The Serbian Chemical Society

WoS Q Değeri

Q4

Scopus Q Değeri

Q3

Cilt

70

Sayı

10

Künye