Flow Injection Amperometric Detection of Sulfide Using a Prussian Blue Modified Glassy Carbon Electrode

[ X ]

Tarih

2012

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Japan Soc Analytical Chemistry

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

This study investigates a new approach for the electrocatalytic determination of sulfide in a flow injection analysis (FIA) system using a Prussian blue modified glassy carbon electrode (PB/GCE). The results from experiments show that PB/GCE significantly enhances the electrocatalytic activity towards sulfide oxidation. A homemade flow electrochemical cell was used to perform the electrocatalytic determination of sulfide in the FIA system. The currents obtained from amperometric measurements in the HA system at optimum conditions (carrier solution, pH 8.0; Britton-Robinson buffer solution containing 0.1 M KCl; flow rate, 1.4 mL/min; transmission tubing length, 10 cm; injection volume, 100 mu L; constant applied potential, +150 mV vs. Ag/AgCl/KClsat) were linearly correlated with the sulfide concentration. A calibration curve was obtained for sulfide concentrations in the range of 0.5 - 100 mu M. The detection limit was found to be 0.3 mu M for the amperometric method. The proposed method was successfully applied to wastewater sample. Finally, results from sulfide measurements by PB/GCE were in good agreement with those obtained from the spectrophotometric method.

Açıklama

Anahtar Kelimeler

Methylene-Blue, Photoelectrocatalytic Oxidation, Electrocatalytic Oxidation, Selective Determination, Ferrocene Sulfonates, Hydrogen-Sulfide, Film, Chromatography, Reduction, System

Kaynak

Analytical Sciences

WoS Q Değeri

Q3

Scopus Q Değeri

Q3

Cilt

28

Sayı

11

Künye