Adsorption and deposition-assisted anodic stripping voltammetry for determination of antimony(III) in presence of hematoxylin on glassy carbon electrode

[ X ]

Tarih

2017

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Elsevier Science Bv

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

A validated simple, reliable and sensitive adsorptive anodic stripping voltammetric procedure is suggested for the accurate determination of antimony(III) on a glassy carbon electrode (GCE) in the presence of hematoxylin (HT), used as a chemical receptor in antimony analysis for the first time. Under optimized conditions, the plot of stripping peak current versus Sb(III) concentration showed two dynamic linear ranges, 1.0x10(-4) - 0.01 mu m (R-2 =0.9986) and 0.01 - 1.0 mu m (R-2 =0.9973) with a detection limit of 9.5x10(-11) mol (11.57 ng L-1). The RSD values of intra-day precision and inter-day precision (n =5) for 5.0x10(-7) mol L-1 Sb(III) were calculated as 1.63% and 2.27%, respectively, indicating that the proposed method has a good precision. Recoveries for Sb(III) in water samples were found to be between 97% and 102% for the 5.0x10(-7) mol L-1 Sb(III), proving the reliability and accuracy of the method proposed.

Açıklama

Anahtar Kelimeler

Hematoxylin, Antimony(IlI) determination, Adsorptive anodic stripping voltamtnetry, Glassy carbon electrode

Kaynak

Talanta

WoS Q Değeri

Q1

Scopus Q Değeri

Q1

Cilt

164

Sayı

Künye