Electroanalytical Investigation of Paracetamol on Glassy Carbon Electrode by Voltammetry

[ X ]

Tarih

2015

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Esg

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

An electroanalytical method was developed for the direct quantitative determination of paracetamol (or acetaminophen) in tablet dosage forms based on its oxidation behavior. The electrochemical oxidation and determination of paracetamol were easily carried out on glassy carbon electrode (GCE) using a variety of voltammetric techniques. The electrochemical measurements were carried out on GCE in various buffer solutions in the pH range from 0.51 to 12.00 by cyclic voltammetry (CV) and differential pulse voltammetry (DPV) techniques. The dependence of pH on the anodic peak current and peak potential was investigated. Acetate buffer (pH 4.51) was selected for analytical purposes. Scan rate studies were also completed. The diffusion controlled nature of the peak was established. A linear calibration curve for DPV analysis was constructed in the paracetamol concentration range from 4x10(-6) mol/L to 1x10(-4) mol/L. Limit of detection (LOD) and limit of quantification (LOQ) were obtained as 3.69x10(-7) mol/L and 1.23x10(-6) mol/L respectively. Validation of applied voltammetric techniques was checked with recovery studies.

Açıklama

Anahtar Kelimeler

Voltammetric determination, drug forms, glassy carbon electrode, paracetamol, voltammetry

Kaynak

International Journal of Electrochemical Science

WoS Q Değeri

Q3

Scopus Q Değeri

Cilt

10

Sayı

2

Künye