Computational Mechanical Analysis of AO 44A1, 44B1, and 44C1 Fractures with Finite Element Modeling: Evaluation of Screw, Plate, and Kirschner Wire Fixation

[ X ]

Tarih

2024

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Amer Podiatric Med Assoc

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

Background: The aim of this study was to create AO 44A1, 44B1, and 44C1 fractures using finite element analysis to determine the stability of Kirschner wire, intramedullary screw, and plate-screw fixation methods in fracture. Methods: Using finite element analysis, the postreduction behavior of AO 44A1, 44B1, and 44C1 fractures with Kirschner wire, intramedullary screw, and plate-screw fixation methods was analyzed and compared in terms of displacement and stress. Results: The lowest amount of displacement was provided with the intramedullary screw method in AO 44A1 and 44B1 fractures and with the 4-mm Kirschner wire method in AO 44C1 fractures. The total displacement of the intramedullary screw system used for fixation in AO 44A1, 44B1, and 44C1 fractures was lower. Conclusions: According to finite element analysis results, the lowest amount of displacement was obtained with intramedullary screw fixation in AO 44A1 and 44B1 fractures, and 4mm Kirschner wire fixation was achieved in AO 44C1 fractures. (J Am Podiatr Med Assoc 114(1), 2024 doi:10.7547/22-155)

Açıklama

Anahtar Kelimeler

Case Series, Fibula, Surgery

Kaynak

Journal of The American Podiatric Medical Association

WoS Q Değeri

N/A

Scopus Q Değeri

Q3

Cilt

114

Sayı

1

Künye