Development of an amperometric hydrazine sensor at a disposable poly(alizarin red S) modified pencil graphite electrode

dc.authoridKarakaya, Serkan/0000-0002-6401-3295
dc.contributor.authorKarakaya, Serkan
dc.date.accessioned2025-01-27T20:32:05Z
dc.date.available2025-01-27T20:32:05Z
dc.date.issued2019
dc.departmentÇanakkale Onsekiz Mart Üniversitesi
dc.description.abstractIn the present study, electrocatalytic oxidation of hydrazine was examined by both cyclic voltammetry and amperometry methods at a disposable, poly(alizarin red S) modified pencil graphite electrode (poly(ALS)/PGE) for the first time. Electrochemical behaviors of both bare PGE and poly(ALS)/PGE were tested by cyclic voltammetry towards 1.0 mM hydrazine in pH 9.0 Britton-Robinson buffer solution. Cyclic voltammograms showed that the oxidation peak potential of hydrazine at PGE (+ 0.70 V) was significantly reduced at poly(ALS)/PGE (+ 0.35 V) and also, a significant increment in oxidation current was observed. The recorded CVs confirmed that poly(ALS) shows high electrocatalytic effect towards oxidation of hydrazine. Secondly, the effects of monomer concentration, cycle number used in electropolymerization step and pH value of supporting solution on oxidation current of hydrazine were also optimized and found to be 1.0 mM, 5 cycle, and pH 10.0, respectively. The designed sensor shows a wide linear response range between 1.0 and 600 mu M hydrazine with a LOD of 0.28 mu M and a sensitivity of 833 mu A mM(-1) cm(-2). The selectivity property of the poly(ALS)/PGE was tested in presence of various interferences. Applicability studies of the poly(ALS)/PGE were also performed on different water samples and high recovery values (95-108%) were obtained. [GRAPHICS]
dc.identifier.doi10.1007/s00706-019-02513-4
dc.identifier.endpage1920
dc.identifier.issn0026-9247
dc.identifier.issn1434-4475
dc.identifier.issue11
dc.identifier.scopus2-s2.0-85074387055
dc.identifier.scopusqualityQ2
dc.identifier.startpage1911
dc.identifier.urihttps://doi.org/10.1007/s00706-019-02513-4
dc.identifier.urihttps://hdl.handle.net/20.500.12428/23349
dc.identifier.volume150
dc.identifier.wosWOS:000492559100001
dc.identifier.wosqualityQ3
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.language.isoen
dc.publisherSpringer Wien
dc.relation.ispartofMonatshefte Fur Chemie
dc.relation.publicationcategoryinfo:eu-repo/semantics/openAccess
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_WoS_20250125
dc.subjectHydrazine
dc.subjectAlizarin red S
dc.subjectPencil graphite electrode
dc.subjectRedox reactions
dc.subjectSensor
dc.subjectElectrochemistry
dc.titleDevelopment of an amperometric hydrazine sensor at a disposable poly(alizarin red S) modified pencil graphite electrode
dc.typeArticle

Dosyalar