İncebeyaz, BurakEren, HakanÖztaş, Bengi2025-01-272025-01-2720222757-6744https://doi.org/10.52037/eads.2022.0009https://search.trdizin.gov.tr/tr/yayin/detay/536028https://hdl.handle.net/20.500.12428/17050Artifacts (error) encountered in phosphor plates (PSP) used in intraoral digital imaging were evaluated under a new classi?cation.\rIn this review, when obtaining images with intraoral phosphor plates, artifacts before, during, after the irradiation and artifacts\rcaused by the scanner were determined according to the review of the literature. Intraoral radiographs are obtained while many\rartifacts are encountered. These artifacts can occur in both conventional and digital systems. Artifacts that occur before irradiation\rsuch as scratches, cracks, ?ngerprints, dust particles, bite marks, plate edge peeling; during irradiation such as positioning error,\rmotion artifact, shooting from different angles errors, cone-cut, double image, light-dark images, glare, fading, revers\rirradiation, refraction, distortion; after irradiation such as irregular image density, noise, writing artifact, ?ngerprints, fading\rartifacts; and depending on the scanner artifacts such as inclined placement, a parallel line to scan direction, dust fragment, plate\rsize determination error has been shown to be visible. An inaccurate image with no diagnostic value requires regeneration. For\rthis reason, knowing the causes of artifacts in radiographic images is important for the patient, the environment, and the\rpractitioner to prevent unnecessary x-rays (radiation).eninfo:eu-repo/semantics/openAccessMikroskopiDiş HekimliğiGörüntüleme Bilimi ve Fotoğraf TeknolojisiRadyolojiNükleer TıpTıbbi GörüntülemeNew Classi?cation of Intraoral Phosphor Plate Artifacts Based on Literature ReviewReview Article491465210.52037/eads.2022.0009536028